Calibrate an Image

To measure a distance on a 2D image, the image must be calibrated first.

If you add a measurement to an uncalibrated 2D image, calibration will automatically start when you click Measure or Segment.

Click a first point and a second point of the reference measurement.

Enter the Reference value of the indicated measurement.

Click Apply.

The calibration measurement is shown on the image and the Calibration object is added to the Visibility smart panel.

Note

If you calibrate an image that was created with an USB image sensor, you can immediately create measurements on all the images that were acquired with this USB intraoral sensor or PSP device.

Recalibrate an Image

To recalibrate an image:

On the title bar of a 2D viewer, click the drop-down arrow.

Select Recalibrate image.

Click a first point and a second point of the reference measurement.

Select a recalibration method:

Reference value: Enter the reference value of the indicated measurement.

Magnification factor (only available for precalibrated images, either by device or DICOM tag info): Enter the magnification factor value.

Click Apply.

Edit a Calibration

Reference Measurement

To edit the value of the calibration measurement:

Open the Visibility smart panel.

Click More options … next to the Calibration object and select Edit calibration.

Enter the new Reference value and click Apply.

To edit the calibration, drag the control points to adjust the position of the calibration object in the viewer. All measurement values are updated accordingly.

To move the label of the calibration measurement, drag it.

Edit Magnification Factor

To edit a previously set magnification factor:

On the title bar of the Clinic module viewer, click the drop-down arrow.

Select Edit magnification factor.