SmartFusion™ Warnings Used

  Dental scan manually aligned.
This option might introduce small inaccuracies. Check the dental scan alignment carefully. If incorrect, make sure the corresponding points are positioned correctly.
  When the user chooses to apply manual alignment for the SmartFusion™ of the dental scan and the patient model, the system will issue a warning to double-check the alignment to avoid possible ininaccuracies.
  One or more pairs of points are not corresponding.   The system measures the distance between the points placed on the patient model and the corresponding points indicated on the dental scan. If the distance is greater than expected, the system issues a warning and requests the user to adjust the pairs of points.
  The defined points are insufficiently distributed over the dental scan.   When the system detects that the defined points are not placed in at least two out of four of the quadrants, a warning is generated. The user is requested to adjust the corresponding points, or to add a pair of corresponding points, taking into account good distribution over the quadrants of the dental arch. Clustering should be avoided.
  Indicate at least three corresponding points.  

Not enough pairs of points have been defined to be able to finish the wizard. Place additional pairs of points until the minimum number of point pairs to be placed is reached. 

 